Sistem fix si stationar de monitorizare calităţii aerului
Canale de masurareTSP, PM10, PM2.5, PM1.0, viteza si directia vantului, debitul ploiele, directia vantului, temperatura si umiditate
Model OSIRIS

Statie de prelevare si monitorizare calitatea aerului înconjurător si a datelor climatice

Portable and stationary direct reading airborne particle sampler


Pret : la cerere


Livrare: 7 - 10 zile lucratoare


produsul poate fi achizitionat si prin SEAP


solicit oferta personalizata »

staţie automată de monitorizare a calităţii aerului,

Domenii de aplicatii

Date tehnice

Model OSIRIS
statie mobila de prelevare particule de aer cu citire directa masurare continua concentratiei particulelor in micro-gram de TSP, PM10, PM2.5 and PM1 cu rezolutie 0.1 microgram / m3
Alimentare cu baterie
Pompa de prelevare integrata
Filtru de referinta pentru calibrare gravimetrica
Intefata (intrare) pentru senzor viteza si directia vantului
Interfata (intrare) pentru numarator de trafic
Interfata (iesire) pentru dispozitiv de incalzire si pentru alarma
Intrari standard TSP (1mm din plasa otel inox)
Intrare (admisie) cu incalzire Heated inlet incalzire pana la 60°C
Detector laser neflometric
Mod aplicatii de mediu TSP, PM10, PM2.5, PM1.0
Mod aplicatii locul de munca Inhalabil, toracic, respirabil
Domeniu de measura 0 la 6000 micrograme / m3
Limita de detectie 0.01 micrograme / m3
Domeniu de monitorizare 0 la 60mg/m3 fara dimensionare particule
Domeniu dimensionarea particulelor 0.5 la 20 diametru micron
Mod numararea particulelor 3 canale particule/cc
Debit 600cc / minut
Filtru de referinta Ø25mm GFA
Temperatura de operare -5°C la +50°C
Securitate protectie cu parola
Alarma Sirena, sms telefon mobil, far vizual si email
Display 2 linii alfanumeric cu iluminare de fundal
Stocare datelor memorie interna cu baterie separata de rezerva (backup) 128k biti
Perioada de calculare a mediei 1 sec. pana la 4 ore
Tip baterie reincarcabila, acid/plumb sigilat interna 6v 2.8 AH
Alimentare cu dispozitiv de sursa externa 80 .. 260v AC, weatherpoof optional
Intrari pentru meteorologice viteza si directia vantului, precipitatii, temperatura si umiditate
Alte intrari de logare 2 intrari analogice 0 la 5 volt
Interfete RS232 I/O 9600 baud via PC-link
Telemetrie I/O 1200 baud opto izolat
Iesiri analogice 0 .. 4 volt analogic de TSP sau canalul de PM10, 12 bit rezolutie optional
Montare Carcasa pe perete sau pe bara cu incuietor
Clasa protectie carcasa IP66
Dimensiune Externa 260 x 160 x 150 mm  
Greutate 11.8kg  
Optiuni pentru alimentare solara, vant, electrica, baterie

NB. Va rugam sa retineti: senzorii si accesoriile descrise in prezentarea produsului de mai sus sunt toate optionale si trebuie sa fie comandate separat.

Main Features
Portable direct reading airborne particle sampler which can be used as a portable instrument or deployed in a semi-permanent installation.
Fitted into a Turnkey Lamp Post Box, Osiris can be quickly deployed to study short or long term pollution hotspots in towns and cities and exceedance areas rapidly determined.
Continuously indicate the concentration of TSP, PM10, PM2.5 and PM1 particles with a resolution of 0.1 micrograms per cubic metre. In its workplace mode, it will indicate inhalable, thoracic and respirable concentrations.
Osiris is Environment Agency's MCERTS certified, ensuring its accuracy in recording data.
An internet-based server is available for Osiris, providing a web interface, camera capability and internet communication

Advantages
Battery or mains operation
Internal sampling pump
Reference filter for gravimetric calibration
Inputs for wind speed and direction and traffic counters
Outputs for inlet heaters and alarms

Technical data

Model OSIRIS
Portable direct reading airborne particle sampler Continuously indicate the concentration of TSP, PM10, PM2.5 and PM1 particles with a resolution of 0.1 micrograms per cubic metre
Battery or mains operation
Internal sampling pump
Reference filter for gravimetric calibration
Inputs for wind speed and direction and traffic counters
Outputs for inlet heaters and alarms
Standard inlet TSP (1mm stainless mesh)
Heated inlet Heating to 60°C
Detector Turnkey laser nephelometer
Environmental mode TSP, PM10, PM2.5, PM1.0
Workplace mode Inhalable, thoracic, respirable
Measurement range 0 to 6000 micrograms per cubic metre
Detection limit 0.01 micrograms per cubic metre
Indicator range 0 to 60mg/m3 without particle sizing
Particle size range 0.5 to 20 micron diameter
Particle counting mode Three size channels in particle per cc
Flow rate 600cc per minute
Reference filter 25mm diameter GFA circle
Operating temperature -5°C to +50°C
Security Password protection
Alarm Siren, text to cellular phone, visual beacon and email
Display Two line alphanumeric with backlight
Data storage Internal with separate battery backup 128k byte
Averaging period 1 second to 4 hours
Battery Sealed lead acid, rechargeable Internal 6v 2.8 AH
Sampling current drain Including heated inlet and backlight 1.2A
External power pack 80 to 260v AC input, weatherpoof optional
Meteorological inputs Wind speed and direction, rainfall, temperature and humidity
Other logging inputs Two 0 to 5 volt analogue inputs
Interfaces RS232 I/O 9600 baud via PC-link
Telemetry I/O 1200 baud opto isolated
Analogue output 0 to 4 volt analogue of TSP or PM10 channel, 12 bit resolution optional
Wall or lamppost box Lockable steel
Case protection To IP66 (excluding inlet and exhaust)
Dimensions External dimensions in mm 260 x 160 x 150
Weight Instrument and enclosure approximate weight in kg 11.8kg
Power options Solar, wind, mains and battery

 

Optionale / Optionals

AirQ environment monitoring software Collect data from Osiris
Searchable database of results
Linear and polar (pollution rose) graphs
Report preparation
AirQWeb web-based dust monitoring Monitor instruments via a web interface
View results and plot graphs
Produce reports
Manage alerts and settings
Web Server Internet connectivity for environmental instruments Allows communication internet communication to the monitors via a fixed line or 3G/4G mobile broadband connection
AlarmBox Dust Suppression Controller on multiple conditions across TSP, PM10, PM2.5, wind etc.
Internet Gas Monitor Records up to 8 gas species simultaneously
Automatic zeroing virtually eliminates drift
Integrated with AirQWeb for live monitoring
Can be used alongside Turnkey dust monitors
Class 2 Noise Monitor Measures LAeq, LAmax, LA10, LA90 and more
Connects to Osiris for continuous dust & noise monitoring
Monitor & record via internet with AirQWeb
IP65 enclosure protects against dust & water
Class 1 Internet Noise Monitor Multichannel noise monitoring
Alarms via email, text message or visual traffic light system
Class 1 performance over a 95 dB dynamic range
Leq, L10, L90 Lmax and more in both dB(A) and dB(C) simultaneously
Internet Vibration & Noise Monitor Multichannel noise & vibration monitoring
Alarms via email, text message or visual traffic light system
Class 1 performance over a 95 dB dynamic range
Leq, L10, L90 Lmax and more in both dB(A) and dB(C) simultaneously
Vibrosound Analyser Simultaneous Vibration & Type 1 Noise Leq monitoring
Automatic unattended monitoring & downloading
Directly export graphs & reports to Excel
Remote data monitoring with SMS text alerts
Anemometer  
Battery Box for 12V Leisure Battery  
Battery Holster for NP5 12 SLA Battery  
GF/A Filters 100-pack  
GSM Modem  
Osiris UK Battery Charger  
Rain Collector  
Temperature and Humidity Sensor  
TSP Head  
US Robotics Modem